Fan - Philips Respironics V60 Service Manual

Respironics v60 ventilator
Hide thumbs Also See for Respironics V60:
Table of Contents

Advertisement

7.6 Fan

Fan filter
Fan filter housing
Fan mount
push pin (x4)
1049766 Rev A
Follow these steps to remove the fan (Figure 7-6). Reverse to install.
NOTE:
See Chapter 4 for fan filter replacement instructions.
1. Remove top cover (section 7.4).
2. Remove fan mount push pins that attach the fan to the back panel
(pliers may be required).
3. Disconnect fan cable from the motor controller (MC) PCBA.
4. Remove the fan from inside the back panel.
5. Remove fan filter assembly (fan filter and housing) from outside the
back panel.
NOTE:
When installing, orient the fan as shown (cable at the top inside of
the fan). Orient the fan filter and housing as shown.
Install fan as shown:
^ AIR FLOW (into ventilator)
< ROTATION
V60 Ventilator Service Manual
Component Removal/Installation
Figure 7-6: Removing the Fan
Chapter 7
Fan cable
(to MC PCBA)
Fan
Fan isolation
mount (x4)
Back panel
7-7