Sartorius FB Operating Instructions Manual page 60

Factory series electronic precision scales
Table of Contents

Advertisement

Printing a Data Record
Purpose
You can generate a printout of
weights, other measured values and
identification numbers for
documentation purposes. You can
format the printout to meet individual
requirements.
60
Available Features
Print manually/automatically:
To print the information contained in
the measured value line (weight
readout, calculated value, numeric
entry, alphabetic entry)
Line format: You can configure
a data ID code of up to 6 characters
for each of the values printed; this
data ID code is printed at the
beginning of the line
Sample ID: You can configure an
extra line for identification of each
weighed or calculated value
Print application parameters:
You can generate a printout of the
values configured for initialization of
an application before printing the
measured results
ISO/GLP-compliant printout:
To print out parameters relating
to weighing conditions
Auto print: To have a printout
generated automatically when certain
conditions are met, e.g., time
elapsed, stability reached, etc.
Print net-total: For a printout of
a component or total weight when
using the "net-total" application
Print animal weights: For an
automatic printout of animal weight,
or of animal weight plus calculated
weight after averaging
Auto print +/–: for automatic
printout of a weight when it lies
within preset limits at stability
Factory Settings
Print manually/automatically:
Printout generated manually
(by pressing p/[PRINT])
or automatically, depending
on stability parameter:
Manual after stability (6 1 2)
Line format:
A data ID code of up
to 6 characters preceding weighed
or calculated values:
For other applications/GLP (7 2 2)
Print application parameters:
Print one or more of the initialization
values for the current application
program: Off (7 1 1)
ISO/GMP-compliant printout/
data record: Documentation of
weighing conditions for each series
of measurements/each lot:
Off (8 10 1)
Auto print:
Automatic printout of weighed/
calculated values:
not a factory setting; see
"Print manually/automatically" (6 1 2)
Stop auto print: not possible
by pressing p/[PRINT] (6 2 2)
After 1 display update (6 3 1)
Print net-total:
Printout of component weight (net
weight) or total weight (tare weight):
Auto print net (7 3 1)
Print animal weights:
Automatic printout of animal weight,
or of animal weight plus calculated
result: On: animal wt. (3 9 2)
Auto print +/–:
Automatic printout of weights when
within tolerance at stability:
off (4 2 2)
Parameter settings:
See "Configuring the Scale"

Hide quick links: