Practical Example
Checkweighing samples of 170 g, with an allowable tolerance of –5 g and +10 g.
Printout of upper and lower tolerance limits. Weighed values printed out automatically when stability is reached
and weight is within the control range.
Step
1. Select the checkweighing
application in Setup
2. Set the following checkweighing
parameters in the Setup menu:
– Auto print +/–: On: values in tol.
– Print app. param.: All parameters
3. Prepare a container for the sample
4. Tare the scale
5. Enter initialization values
6. Store target value (here: 170 g)
7. Store target value and
unload scale
8. Enter value for lower limit
(170 g – 5 g)
Key (or instruction)
see "Preparation"
See "Configuring the Scale"
Place empty container
on the scale
t
start soft key
v/[FUNCTION]
Place ideal sample
in container
setp soft key
v/[FUNCTION]
Remove ideal sample
from scale
1 6 5