Internal Calibration/Adjustment
The menu code setting 1 9 4 must be
selected in the Setup menu.
Inside the scale housing is a built-in,
motorized calibration weight.
The internal calibration/adjustment
sequence is as follows:
Select the calibration function:
Press [ISOTEST]/[TEST]
> The internal calibration weight
is applied automatically
> The scale is calibrated
> If the setting for "Calibration
automatically followed by
adjustment" (1 10 1) has been
selected in the Setup menu, the
scale will now be automatically
adjusted
> The internal calibration weight
is removed
28
Calibration and Adjustment Sequence
In the Setup menu, you can configure
the scale so that:
– calibration is always followed
automatically by adjustment
(cal. -> auto adjust, 1 10 1), or
– you have the choice of ending
the sequence or starting adjust-
ment after calibration
(cal. -> man. adjust, 1 10 2).
If no deviation is determined during
calibration, or the deviation is within
the tolerance limits dictated by the
degree of accuracy you require,
it is not necessary to adjust the scale.
In this case, you can end the
calibration/adjustment sequence
after calibration. There are 2 soft
keys active at this point:
– ext. adjust or int. adjust
to start adjustment
– end to end the sequence