8 Basic Operation

8.1 Switching on/off

To switch-on actuate forward the switch-on/switch-out on the right lower side of
the balance. The balance will carry out a self-test The balance is ready for
weighing when the weight display appears.
To switch-off push backward the switch-on/switch-out on the right lower side of
the balance.

8.2 Zeroing

Resetting to zero corrects the influence of light soiling on the weighing plate. The
resetting range of the balance was adjusted to ± 2% max. by factory.
Further adjustments are possible in the menu, see chapter 13.
When using as counting system the zeroing range of both balances can be set in
the menu, see chapter 13.
Manual
Unload the balance
Press
, the balance starts resetting to zero.
The [ ] symbol over a appears.
Automatic
In the menu the automatic zero point correction can be switched off or the amount
can be changed, see chapter 13.

8.3 Switch over balance/weighing bridge

For piece counting a weighing bridge can be connected via the second balance in-
terface. In the counting system KERN CCS the quantity counting is made on the
bulk material balance KERN KFP. The reference balance KERN CFS permits due
to its high resolution a very precise determination of the average piece weight.
The second balance can be operated in the same manner like the first balance.
20