7 Adjustment

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory).
This adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is re-
quired for stabilization. Ensure that there are no objects on the weighing plate.
• Info about adjustment weights can be found on the Internet at:
http://www.kern-sohn.com
• In order to avoid errors at the piece quantity determination, both balances
must be adjusted with the same acceleration due to gravity.
In case of non-compliance counting errors will result!

7.1 Adjustment models KERN CFS/CCS

Switch-on balance and during the self test press
.
Use the number keys to enter password:
Either
• Standard password „0000"
or
• personal password, input see chapter 13
Confirm input with
When using as counting system, the bulk material
balance as well as the reference balance must be
adjusted. The adjustment process must be carried
out on both balances.
Select bulk material or reference balance via
The appeared [ ] displays the current balance.
Acknowledge with