7.2 Adjustment bulk weighing balance, not preconfigured by KERN

Switch-on balance and during the self test press
.
Use the number keys to enter password:
Either
• Standard password „0000"
or
• personal password, input see chapter 13
Confirm input with
Select bulk material balance using
The appeared [ ] displays the current balance.
Acknowledge with
If necessary, at balance zero display using
select the weighing unit [kg or lb], which shall be
used for adjustment. The appeared [ ] displays the
current weighing unit.
Acknowledge with
Ensure that there are no objects on the weighing
plate. After successful stop check press the
ton.
Enter value of the necessary adjustment weight
using numeric keyboard and acknowledge with
In order to achieve high-quality weighing results in
the sense of the measuring technology, it is recom-
mended to select the nominal value as high as pos-
sible. We recommend 80 % max.
Place the required adjusting weight carefully in the
centre of the weighing plate and press
firm.
After the adjustment the balance will carry out a self-
test. Remove adjusting weight during self test,
balance will return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message
(
CFS/CCS-BA-e-1013
Operation
.
.
.
.
), repeat adjustment process.
but-
.
to con-
Display