10 Counting

During piece counting parts can either be counted into a container or out of a con-
tainer. To count a greater number of parts the average weight per part has to be
determined with a small quantity (reference quantity). The larger the reference
quantity, the higher the counting exactness.
High reference must be selected for small parts or parts with considerably different
sizes.
The average piece weight can only be determined by stable
weighing values.
If weighing values are under zero, the piece counter display shows a
negative number of items.
The exactness of the average piece weight can be increased at any time
during piece counting, by entering the displayed number of items and con-
firming by
the additional pieces increase the base for the calculation, the reference
also becomes more exact.

10.1 Determination of the average piece weight by weighing

Set reference
Reset balance to zero or tare the empty weighing container if necessary.
Place on the weighing plate a known number (e.g. 10 items) of individual pieces
as a reference.
Wait for the stability display, than enter the number of individual items via the nu-
meric keypad.
Confirm within 5 sec by pressing
The balance determines the average piece weight.
Count the items
Tare if necessary, place weighing good and read off the number of items.
When an optional printer is connected, the displayed value can be edited by
pressing
(menu settings „P mode"
26
. After the reference optimization sounds a signal tone. As
.
„Print"
„Au OFF", chpt. 12.2).