10.2 Numeric input of the average piece weight

Set reference
Enter the average piece weight via the numeric keyboard and acknowledge within
5 sec using button
If in the weight display as weighing unit [kg] is active, the average piece weight
will be displayed in [g]. If as weighing unit [lb] is active, the average piece weight
is also displayed in [lb].
Count the items
Tare if necessary, place weighing good and read off the number of items.
When an optional printer is connected, the displayed value can be edited by
pressing
(menu settings „P mode"

10.3 Automatic reference optimization

If at the reference determination the placed weight or the placed piece number is
too small, in the display of the average piece weight the triangle symbol will appear
[
]
[
]
over
or
To optimize the calculated average piece weight automatically, add further parts
whose number is smaller than that of the first reference determination. After the ref-
erence optimization sounds a signal tone. For each reference optimization the aver-
age piece weight is newly calculated. As the additional pieces increase the base for
the calculation, the reference also becomes more exact.
By pressing the
key, recalculation can be avoided and the reference weight can
be blocked.
Automatic reference optimization will be deactivated as soon as the number of
added parts exceeds the saved reference quantity.
CFS/CCS-BA-e-1013
.
.
„Print"
„Au OFF", chpt. 12.2).