9 Basic Operation

9.1 Turn on/off

Start-up:
Internal calibration
The internal calibration must be performed in the following cases:
-
Each time the balance is switched on and disconnected from the power
supply.
-
Each time the balance is switched on in battery or rechargeable battery
mode.
-
After the adjustment interval has been reached, see chap 7.9.2.
Switching off:

9.2 Simple Weighing

24
Press the ON/OFF button.
The display lights up and the balance carries out an self-
test.
If the balance has been disconnected from the mains
voltage, an internal calibration must be carried out, see
Chap.7.9.1.
Wait until the weight display appears
The scales are now ready to weigh using the last active
application
Keep ON/OFF button pressed until the display disap-
pears
Check zero display [>0<] and set to zero with the help of
the ZERO key, as required.
Place goods to be weighed on balance
Wait until the stability display appears (
Read weighing result.
).
TPCJ-BA-e-2310