17.1 Error messages

Error message
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.
46
Description
Zeroing range exceeded
due to switching-on balance
or pressing
(normally
4% max)
Keyboard error
Value outside the A/D
changer range
Adjustment error
Possible causes
• Object on the weighing
plate
• Overload when zeroing
• Improper adjustment
• Damaged weighing cell
• Damaged electronics
• Improper operation of
the balance
• Weighing plate not in-
stalled
• Damaged weighing cell
• Damaged electronics
• Improper adjustment
CFS/CCS-BA-e-1013