Explanation Of Operating Keys - YOKOGAWA SC202G (S) User Manual

Conductivity and resistivity transmitter
Table of Contents

Advertisement

4-2 Operation
Output hold flag
Main display
Message display
Key prompt flags
Selection keys
YES : Accept setting
NO
: Change setting
Adjustment keys
>
: Choose digit to
adjust
^
: Adjust digit
ENT : Confirm change
Figure 4-1. SC202 operator interface

4-2. Explanation of operating keys

MODE key
This key toggles between the measuring and maintenance modes. Press once to obtain
access to the maintenance function menu.
Press again to return to the measuring mode (press twice when hold is activated).
YES/NO keys These are used to select choices from the menu.
YES is used to accept a menu selection.
NO is used to reject a selection, or to move ahead to the next option.
DATA ENTRY keys (
is used as a "cursor" key. Each press on this key moves the cursor or flashing digit
one place to the right. This is used to select the digit to be changed when entering
numerical data.
is used to change the value of a selected digit. Each press on this key increases the
value by one unit. The value can not be decreased, so in order to obtain a lower
value, increase past nine to zero, then increase to the required number.
When the required value has been set using the > and ^ keys, press ENT to confirm
the data entry. Please note that the EXA does not register any change of data until
the ENT key is pressed.
key
This is the commissioning mode key. It is used to obtain access to the commissioning
*
menu. This can only be done with the cover removed or opened. Once this button has
been used to initiate the commissioning menu, follow the prompts and use the other keys
as described above.
IM 12D7B3-E-E
HOLD
YES
CALIB
DISP 1
DISP 2 - (Only when second temp. compensation enabled)
HOLD - (only when enabled)
mA
)
Fail flag
MODE
FAIL
MEASURE
CAL
DISPLAY
YES NO
HOLD
ENT
NO
MODE
ENT
Broken line indicates area
that can be seen through
front cover
Menu pointer flags
Commissioning
function menu
OUTPUT
SET HOLD
TEMP.
SERVICE
Commissioning
mode access key
Measure/Maintenance
mode key