6-3 Calibration

6-3. Calibration with HOLD active

YES
NO
MODE
MODE
HOLD
YES
NO
YES
HOLD
YES
NO
Put the sensor in standard
solution. Press YES.
Set the value
using the >, , ENT key.
Select the flashing digit with the > key.
Increase its value by pressing the key
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press YES.
HOLD will be displayed. Press NO to turn off
HOLD and return to the measuring mode.
IM 12D7B3-E-E
MODE
MEASURE
CAL
DISPLAY
HOLD
ENT
HOLD
ENT
appears, and the YES/NO
HOLD
ENT
Press the MODE key.
The legend CALIB
key prompt flags flash.
HOLD
ENT
HOLD
YES
NO
HOLD
YES
NO