General Specifications - YOKOGAWA SC202G (S) User Manual

Conductivity and resistivity transmitter
Table of Contents

Advertisement

2-1 Specifications

2. GENERAL SPECIFICATIONS

2-1. Specifications
A. Input specifications : Two or four electrodes measurement
B. Detection method : Frequency, read-pulse position and
C. Input ranges
- Conductivity
Minimum
Maximum
- Resistivity
Minimum
Maximum
- Temperature
Pt1000
Pt100 and Ni100 : -20 to +200 °C (0 - 400 °F)
8K55 NTC
Pb36 NTC
mA
D. Output Span
- Conductivity
- Resistivity
- Temperature
Sensor type
Pt1000
Pt100, Ni100
Pb36 NTC, 8k55 NTC
mA
E. Transmission Signal
F. Temperature compensation
- Reference temp.
G. Compensation algorithm
-NaCl
-T.C.
IM 12D7B3-E-E
with square wave excitation. Cell
constants from 0.008 to 50 cm-1
WU40 sensor cable up to 20m. Up
to 60m total using BA10 junction box
and WF10 extension cable
reference voltage are dynamically
optimized.
: 0.000 µS/cm to 1999 mS/cm at
25 °C (77 °F) reference temperature.
: 0.2 µS x C at process temperature
(underrange 0.000 µS/cm).
: 500 mS x C at process temperature
(overrange 550 mS x C).
: 0.000 ∧U - 999 ] U/C at 25 °C
(77 °F) reference temperature.
: 0.002 ∧U/C at process temperature
(underrange 0.000 kΩ x cm).
: 5 ] U/C at process temperature
(overrange 999 MΩ x cm).
: -20 to +250 °C (0 - 500 °F)
: -10 to +120 °C (10 - 250 °F)
: -20 to +120 °C (0 - 250 °F)
: - min 0.01µS/cm
: - max. 1999 mS/cm. (max 90% zero
suppression)
: - min 0.001kΩxcm
: - max. 999 ] U x cm. (max 90% zero
suppression)
: Dependent on temp. sensor type:
min.
max.
25 °C (50 °F) 250 °C (500 °F)
25 °C (50 °F) 200 °C (400 °F)
25 °C (50 °F) 100 °C (200 °F)
The instrument is user programmable
for linear or non-linear conductivity
ranges.
: Isolated output of 4-20 mA DC .
Maximum load 425 Ω.
Burn up (22 mA) or Burn down
(3.9 mA) or pulse of 22mA to signal
failure. See Fig.2-1 and 2-2.
: Automatic, for temperature ranges
mentioned under C (inputs).
: programmable from 0 to 100 °C or
30 - 210 °F (default 25 °C).
: According IEC 746-3 NaCl tables
(default).
: Two independent user programmable
temperature coefficients, from -0.00%
to 3.50% per °C (°F) by adjustment or
calibration.
- Matrix :
: Conductivity function of concen-
tration and temperature. Choice out
of
5 preprogrammed matrixes and a
25-point user-programmable matrix.
H. Serial Communication
: Bi-directional according to HART
digital communication super imposed
mA
on the 4-20mA signal.
I.
Logbook
: Software record of important events
and diagnostic data. Available through
HART interface.
J. Display
: Custom liquid crystal display, with a
main display of 3
high. Message display of 6 alpha-
numeric characters, 7 mm high.
Warning flags and units (mS/cm,
kΩ.cm, µS/cm and MΩ.cm) as
appropriate.
K. Power supply
: Nominal 24 volt DC loop powered
system.
mA
SC202G
; up to 40 volts
SC202S
: up to 31.5 volts
Note:
The transmitter contains a switched
power supply. The transmitter requires
a minimum Power voltage in order to
work correctly, which is dependant on
the load. Please refer to figures 2-1
and 2-2 for the correct power supply.
1200.0
4 mA
1000.0
800.0
775.0
600.0
425.0
400.0
200.0
0.0
12 14 16 18 20 22 24 26 28 30 32 34 36 38 40
Fig. 2-1. Supply voltage/ load diagram
17 Volts
14.5 Volts
4 mA
7 mA
Output Current (mA)
Fig. 2-2. Minimum terminal voltage at the SC202
/
digits 12.5 mm
1
2
22 mA
1100.0
230.0
31.5 V
(limit for IS version)
Voltage (V)
20 mA