Communication Setup; General; Test And Setup Mode - YOKOGAWA SC202G (S) User Manual

Conductivity and resistivity transmitter
Table of Contents

Advertisement

5-18 Parameter setting

5-7. Communication setup

mA
Code 60
*COMM.
*ADDR.
mA
Code 61
*HOUR
*MINUT
*SECND
*YEAR
*MONTH
*DAY
Code 62
*ERASE

5-8. General

Code 70
*LOAD

5-9. Test and setup mode

Code 80
*TEST
NOTE
: Attempting to change data in service code, 80 and above without the proper instructions
and equipment, can result in corruption of the instrument setup, and will impair the
performance of the unit.
IM 12D7B3-E-H
The settings should be adjusted to suit the communicating device connected to
the output. The communication can be set to HART or to PH201*B distributor (for
Japanese market only).
Select address 00 for point to point communication with 4-20mA transmission.
Addresses 01 to 15 are used in multi-drop configuration (fixed 4mA output).
The clock/calendar for the logbook is set for current date and time as reference.
Erase logbook function to clear the recorded data for a fresh start. This may be
desirable when re-commissioning an instrument that has been out of service for a
while.
The load defaults code allows the instrument to be returned to the default set up
with a single operation. This can be useful when wanting to change from one
application to another.
The test mode is used to confirm the instrument setup. It is based on the
factory setup procedure and can be used to check the QIC (factory generated
Certificate). This test is described in the Quality Inspection Standard, see
chapter 12.