3-1 Installation and wiring

3. INSTALLATION AND WIRING

3-1. Installation and dimensions

3-1-1. Installation site

The EXA transmitter is weatherproof and can be installed inside or outside. It should, however, be installed
as close as possible to the sensor to avoid long cable runs between sensor and transmitter. In any case,
the cable length should not exceed 60 meters (200 feet). Select an installation site where:
Mechanical vibrations and shocks are negligible
No relay/power switches are in the direct environment
Access is possible to the cable glands (see figure 3-1)
The transmitter is not mounted in direct sunlight or severe weather conditions
Maintenance procedures are possible (avoiding corrosive environments)
The ambient temperature and humidity of the installation environment must be within the limits of the
instrument specifications. (See chapter 2).

3-1-2. Mounting methods

Refer to figures 3-2 and 3-3. Note that the EXA transmitter has universal mounting capabilities:
Panel mounting using two (2) self-tapping screws
Surface mounting on a plate (using bolts from the back)
Wall mounting on a bracket (for example, on a solid wall)
Pipe mounting using a bracket on a horizontal or vertical pipe (maximum pipe diameter 50 mm)
162 (6.4)
1/2" SUPPLY
56±0.2
(2.20)
Fig. 3-1. Housing dimensions and layout of
glands
IM 12D7B3-E-E
92 (3.6)
115 (4.5)
1/2" INPUT
M6 bolts (2x)
min. 203
(min. 8.0)
30 (1.2)
CUT-OUT DIMENSION
Fig. 3-2. Panel mounting diagram
154
(6.06)
30 (1.18)
2x ø4
(0.16)
SPACING PANEL
CUT-OUT DIMENSIONS