Operation; Display Functions And Setting; Operator Interface - YOKOGAWA SC202G (S) User Manual

Conductivity and resistivity transmitter
Table of Contents

Advertisement

4. OPERATION; DISPLAY FUNCTIONS AND SETTING

4-1. Operator interface

This section provides an overview of the operation of the EXA operator interface. The basic procedures
for obtaining access to the three levels of operation are described briefly. For a step-by-step guide to data
entry, refer to the relevant section of this user's manual. Figure 4-1 shows the EXA operator interface.
LEVEL 1: Maintenance
These functions are accessible by pushbutton through a flexible front cover window. The functions make up
the normal day-to-day operations that an operator may be required to complete. Adjustment of the display
and routine calibration are among the features accessible in this way. (See table 4-1).
LEVEL 2: Commissioning
A second menu is exposed when the EXA front cover is removed and the display board is revealed. Users
gain access to this menu by pressing the button marked
is used to set such values as the output ranges and hold features. It also gives access to the service menu.
(See table 4-1).
LEVEL 3: Service
For more advanced configuration selections, press the button marked
you reach SERVICE. Now push the "YES" button. Selecting and entering "Service Code" numbers in the
commissioning menu provide access to the more advanced functions. An explanation of the Service Codes
is listed in chapter 5 and an overview table is shown in chapter 11.
Table 4-1. Operations overview
Maintenance
Commissioning
mA
Service
(Access to coded entries
from the commissioning
level)
NOTE:
All three levels may be separately protected by a password. See Service Code 52 in chapter 5 Service
Code table for details on setting passwords.
Routine
Function
CALIB
Calibration with a standard solution or sample
DISPLAY 1&2
Read auxiliary data or set message display
HOLD
Switch hold on/off (when activated)
OUTPUT
Adjust the output range
SET HOLD
Activate the hold function
TEMP 1 & 2
Select method of temperature compensation
SERVICE
Fine tune the specialized functions of the
transmitter
in the lower right of the display board. This menu
*
, then press "NO" repeatedly until
*
Operation 4-1
Chapter
6
4
5
5
5
5
5
IM 12D7B3-E-E

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Exa sc202gExa sc202s