Disposal of Product Waste
As you use the PRO Monitor, you will accumulate solid wastes
that require proper disposal or recycling. These include
batteries, patient applied parts, and packaging material.
.
A leak test of the BP parameter should be performed at least
once
a year or when there is doubt about the validity of the
pressure readings.
Caution: Refer calibration and leak testing to qualified
service personnel. Full calibration details are available in the
.
+5 V
output
on the host port connector.
Calibration
Calibration of the BP parameter should be checked at least
once a year or when there is doubt about the validity of the
readings.
Leak Testing
l
The expected lifetime of the battery largely depends on
the way in which the Monitor is
used. If the battery is
allowed to completely discharge before being fully
recharged, the battery should survive around 200
recharge cycles. If the battery is used in such a way that
it never becomes more than one third discharged and is
fully recharged whenever possible, it can survive up to
1200 cycles. This means that by thoughtful usage, the
lifetime
of the
battery can be extended up to six times.
Replacement batteries may be obtained from Critikon.
Note: The replacement part number of the battery is 63313.
Do not use other types.
Fuses
The Monitor contains five fuses. Two AC line input fuses are
mounted internally and are replaceable only by qualified
service personnel. The remaining three fuses are auto-
resetable and mounted within the Monitor. These fuses
protect the low voltage DC input, the battery, and the