Alarms Button - Critikon DlNAMAP PRO Series 100-400 Operation Manual

Dinamap pro series 100-400
Table of Contents

Advertisement

SpO,. The values and ranges
for these parameters are
not stored when the Monitor is turned off. The
user may
edit the limits, but they are restored to the default values
each time the Monitor is switched on. To permanently
change the alarm limits, refer to "Alarms" under "Service
Button" in the "Using the Menu System" section.
5 minutes.
Note:
Alarm limits are disabled while in Stat mode.
Main
Selection of this button returns the user to the Main menu.
Alarms Button
Selection of this button displays the Alarms menu. This
menu is used to adjust the violation limits for
BP,
Pulse Rate,
and
stat
Selection of this option allows the user to start Stat
determinations. When Stat is selected, blood pressure is
determined as many times as possible in
mmHg, then that setting
is the target pressure that will be used. When the target
pressure is changed, the next determination will use the new
target inflation value if no systolic is available. Initial target
pressure is restored to the factory default
setting after
power-off.
The target pressure can be adjusted permanently
in the Clinician menu of the Service mode (refer to "Press"
in the "Using the Menu System" section).
mmHg. If the target
pressure is set between 100 and 140
mmHg under
the Set BP or Clinical menu, the Monitor automatically
defaults to a target pressure of 110
mmHg for adults and 110 for
neonates. (This is indicated by "AUTO" at the end of
adjustable range.) When using a neonate blood pressure cuff,
if the target pressure is set to greater than
140
mmHg increments. The
factory default is 160
mmHg in 5
Tgt Pressure
Selection of
this option allows the user to set the
BP target
inflation pressure.
The initial target pressure can be set
between 100 and 250
B Using the Menu System