Critikon DlNAMAP PRO Series 100-400 Operation Manual page 10

Dinamap pro series 100-400
Table of Contents

Advertisement

9 For continued safety, use only a power cord of listed type
SJT, three-conductor, min. No. 18 AWG, terminated in a
medical/hospital grade attachment plug, provided with
the following cord tag:
"Hospital Grade Plug." Grounding
14
200
and 500 charge/discharge cycles. When it is necessary to
replace the battery, refer to the "Compatibility Table and
Reorder Codes" listed in Appendix D. To ensure full
charge cycles, replace only with a recommended battery.
If the Monitor is to be stored
for some time, first charge
the battery and then remove it and store it separately from
the Monitor.
BAlTERY IS TOO LOW FOR
MONITOR TO FUNCTION. TURN MONITOR OFF
appears.
The Monitor shuts down all functions until it is turned off and
the battery is recharged or replaced. To run the Monitor on
AC power, it must be powered off and then on again.
Battery charging will take place as long as the Monitor remains
connected to an external AC power source. A battery that is
fully discharged can be fully recharged in
1
hour 50 minutes
when the Monitor is switched off or 8 hours if the Monitor is
switched on.
Notes
l
To prolong the life of the battery, keep the Monitor
connected to an AC outlet whenever possible. NEVER
allow the battery to become completely discharged. A
fully charged battery will power the Monitor for
approximately 2 hours and should survive between
.,:
During battery-only operation, the yellow battery power
indicator LED (17) will light. When the battery becomes
discharged beyond the low battery threshold, the indicator will
begin to flash, and the Monitor will sound warning beeps
every 30 seconds. At this point, the Monitor should be
connected to an AC outlet to recharge the battery. If the
Monitor continues to be used without charging the battery,
the message WARNING: THE
set).
When the Monitor is running on battery
power, a battery
icon
appears in
LCD area 3 (toggling
with the time indicator)
indicating the charge status.
power indicator LED will flash (short flash approx. every 4