Manual Mode; Auto Mode - Critikon DlNAMAP PRO Series 100-400 Operation Manual

Dinamap pro series 100-400
Table of Contents

Advertisement

&Z?)
Auto
BP determinations are started by selecting the AUTO
BP key (27) or the Auto button under the Set BP button
(UK: BP Mode) in the Main menu.
Selecting the AUTO BP key (27) brings up the Set BP menu
(UK: BP Mode) and automatically starts an Auto BP
determination as long as the Monitor is in Manual BP mode.
If the Monitor is already in Auto BP mode, selection of the
AUTO BP key (27) brings up the Set BP menu (UK: BP
Mode) without starting a new determination until the preset
time interval has expired. Pressing the START/STOP
BP key
during
a series of Auto BP determinations will cancel the
determination in progress.
When Auto mode is selected, a number at the right of the
Auto button indicates the time interval between each
reading. To change the time interval, choose the box around
the number and turn the rotor until the desired interval is
reached. The interval can be set between
1
and 120
minutes (1, 2, 3, 4, 5,
10,
15, 20, 25, 30, 45, 60, 90, and
29
mmHg (adult) before another
determination can be started.
BP information will be
displayed
for 2 minutes on the LED unless another
determination is started within that time frame. This applies
to Manual and Vitals (UK:
All
Obs) modes.
After power-off,
the operating
mode returns to the default setting of Manual.
The default
setting of Manual can be overridden to return to
the previous user-selected setting (Auto or Manual) by
selecting Set
BP
(UK: BP Mode) under the Service menu.
Note: The START/STOP BP key is an on-off switch; pressing
it will stop any BP determination (Manual, Auto, Stat, or
Vitals) that is in progress.
Auto Mode
mmHg
(neonate) or 15
@
To start a determination, press the START/STOP BP key (29).
A
normal, uninterrupted Manual cycle takes about 40
seconds.
The cuff pressure must drop below 5
Using the Monitor
Manual Mode