Procedures - Critikon DlNAMAP PRO Series 100-400 Operation Manual

Dinamap pro series 100-400
Table of Contents

Advertisement

mmHg from values for
every inch (2.54 cm) below heart level.
Select appropriate cuff size. Measure patient's limb and
select appropriately sized cuff according to size marked on
cuff or cuff packaging. When cuff sizes overlap
for a
specified circumference, choose the larger size cuff.
Precaution: Accuracy depends on use of proper size cuff.
mmHg to values for every inch (2.54
cm) above heart level. Subtract
1.80
f
Adult/Pediatric
Neonate
3.
4.
If patient is standing, sitting, or inclined, ensure that cuffed
limb is supported to maintain cuff at level of patient's
heart. If cuff is not at heart level, the difference in systolic
and diastolic values due to hydrostatic effect must be
considered. Add
1.80
I
compromised.
Procedures
Using the Monitor
1.
2.
Connect the
end of the air hose which has quick-release
clips to the cuff connector (30) on the front of the Monitor.
Make
sure that the hose is not kinked or compressed.
Note: To
disconnect the hose from the Monitor, squeeze
the quick-release clips together and pull the plug from the
cuff connector (30).
Select the appropriate blood pressure measurement site.
Because normative values are generally based on this site
and as a matter of convenience, the upper arm is preferred.
When upper arm size or shape, the patient's clinical
condition, or other factors prohibit use of the upper arm,
the clinician must plan patient care accordingly, taking into
account the patient's cardiovascular status and the effect of
an alternative site on blood pressure values, proper cuff
size, and comfort. The figure shows the recommended sites
for placing cuffs.
Warning: Do not place the cuff on a limb being used
for intravenous infusion or any area where circulation
is compromised or has the potential to be