SpOz function to find a pulse pattern.
l
An electrosurgical unit (ESU) may be interfering
with
performance.
SOLUTION:
Check the patient.
l
If possible, keep the patient still; check whether the sensor is
applied securely and properly, and replace it if necessary;
move the sensor to a new site; use a sensor that tolerates more
motion.
.
Excessive patient motion may be making it impossible for the
SpO2 value or the pulse rate changes rapidly; the
puise amplitude bar is erratic.
CAUSE:
SpOz
sensor is applied securely and properly, and replace it if
necessary; move the sensor to a new site; or use a disposable
adhesive sensor that may tolerate more motion.
l
Replace the sensor.
PROBLEM: The
SpOz
function to measure saturation and pulse rate.
SOLUTION:
Check the patient.
l
If possible, keep the patient still; check whether the
SpOz function to find a pulse pattern.
l
The sensor may be damaged.
it impossible for the
l
The patient's perfusion may be too low to allow the
Using the Monitor
Troubleshooting
This section discusses potential difficulties and suggestions
for
resolving them. If the difficulty persists, contact a qualified service
person or your local Critikon representative.
The service manual, which is for use by qualified service personnel
provides additional troubleshooting information.
PROBLEM: The pulse amplitude bar indicates a pulse, but no
oxygen saturation or pulse rate values appear on the screen.
CAUSE:
l
Excessive patient motion may be making