SpOZ
measurement.
General Cautions
l
As with any clip-on sensor, pressure is exerted. The
clinician should be cautious in using a clip-on sensor on
SpOZ function is calibrated to read functional
arterial oxygen saturation. Significant levels of
dysfunctional hemoglobins such as carboxyhemoglobin
or methemoglobin may affect the accuracy of the
SpOZ measurement.
l
The
SpOZ measurements.
Always remove oximetry devices and attachments from
the MRI
environment before scanning a patient.
l
The use of cardio-green and other intravascular dyes at
certain concentrations may affect the accuracy of the
SpOZ function during magnetic
resonance imaging (MRI). Adverse reactions include
potential burns to patients as a result of contact with
attachments heated by the MRI radio frequency pulse,
potential degradation of the magnetic resonance image,
and potential reduced accuracy of
SpOz reporting of values when several consecutive
good pulse determinations are made.
Genera/ Warnings
l
Do not use the
SpOz monitoring, the values in the LED flash,
alternating patient values with dashes. The Monitor returns to
normal
ICancel
If the Monitor is unable to detect a pulse for 10 seconds
during normal
SpOz is suspending
2:oo
Sp02.
Low
SpOZ alarm suspension and
return to monitoring
Sp02 alarm suspend is counting down.
Selecting Cancel
will cancel the
SpOz
LED while the
Sp02 is
suspending appears
in Area 2 and dashes appear in the
SpOz
monitoring. A message informing
the user that
SpOz alarm is suspended
for 2 minutes and
then the
PRO returns to normal