Specifications

According to IEC Standard 601-2-7

1 August 2006

Manufacturer

Kodak -Trophy
A subsidiary of Eastman Kodak Company
4, rue F. Pelloutier - Croissy-Beaubourg
77437 Marne-la Vallée Cedex 2
France
Models
• Dental X-ray diagnosis devices, class 1, type B, intermittent use
• KODAK 2200-TR: equipped with tube TRX 708 from
KODAK-TROPHY
• KODAK 2200-C: equipped with tube OCX / 65-G from CEI
Electric power supply (during exposure)
• 230–240 V AC (± 10%), 50 Hz, 5 A, apparent resistance 0.5 Ω
• 100–110–130 V AC (± 10%), 50/60 Hz, 12 A, apparent resistance 0.2 Ω
Electric power supply (no exposure)
• 230–240 V AC (± 10%), 50 Hz, 100 mA
• 100–110–130 V AC (± 10%), 50/60 Hz, 100 mA
Rated high voltage and maximum corresponding current
• Film mode: 70 kV, 7 mA
• RVG mode: 70 kV, 4 mA
Current/voltage combinations for a maximum output power of:
• 490 W in film mode, 70 kV / 7 mA
• 280 W in RVG mode, 70 kV / 4 mA
Rated power for exposure time of 0.1 sec.
• Film mode: 490 W
• RVG mode: 280 W
Rate of use
• At 70 kV, 7 mA and 0.1 sec. and at the maximum tank temperature: one
exposure every 8 sec.
Minimum value of the current/time product in the range of
conformity
• 0.07 mAs at 7 mA
• 0.04 mAs at 4 mA
SM701_K2200_2_en
5-1