Equipped X-ray Generator

5-4
Table 15. Equipped x-ray generator
IEC standard 60601-2-28 (1993)
Type of protection against electric shocks
Degree of protection against electric shocks
Rated value of inherent filtration
Rated value of additional filtration
Rated value of total filtration
Beam-limiting cone, focal spot/skin distance
Maximum accumulated heat
Maximum continuous thermal dissipation
Amount of leaking radiation at maximum rate
during one hour of use
Maximum field of symmetrical radiation
Position and tolerances of the focal point on the
reference axis
Figure 16. X-ray generator
Reference axis
Target angle
Focal point
SM701_K2200_2_en
Conform
Class I
Type B
1.5 mm (0.059 in.) eq. Al
1.0 mm (0.039 in.) eq. Al
2.5 mm (0.098 in.) eq. Al
20 cm (7 7/8 in.)
32,500 J
7 W
< 0.25 mGy
6 cm (2 3/8 in.) diameter
0 mm +/-0.5 mm (0.020 in.)
1 August 2006