Care And Maintenance General Maintenance - Kodak 2200 User Manual

Intraoral x-ray system
Hide thumbs Also See for 2200:
Table of Contents

Advertisement

4
Care and Maintenance
General Maintenance
Important
If the result of any of these checks is not
satisfactory, discontinue using the
equipment and contact an authorized
Kodak Dental Systems technician.
1 August 2006
To make sure that the system functions correctly, you must have it serviced
annually by an authorized KODAK Dental Systems dealer. In addition,
every three months inspect the equipment and make sure of the following:
Generator
• The certification label is legible.
• There are no oil leaks.
Mechanical support
• The wall framework is securely attached to the wall.
• All the labels are legible.
• The scissor arm is stable in all positions.
Control unit and electrical installation
• The symbols are legible.
• The control unit cable and the power supply cable are in good condition.
• The ground is correctly installed.
• The radiology control key returns to its initial position after use.
Functioning
• The audible signal is audible and the x-ray emission light is visible when
you make an exposure (Occlusal, film type 9, 60 kV, 7 mA).
• The message "OP. ERROR" is displayed when you make an exposure
(Occlusal, film type 9, 60 kV, 7 mA) and release the control button
before the exposure time has elapsed.
Timer self-test
• Turn off the system.
• Press and hold the RVG key and turn on the system. When the first light
comes on, release the RVG key.
All the control unit functions and indicator lights will be tested one by
one. At the end of the test, the total number of exposures made by the
machine is displayed.
• A short beep indicates that the test has ended.
SM701_K2200_2_en
4-1