3-7. 1 Volt Range Calibration - Fluke 871A Preliminary Operating Instructions

All solid-state dc differential voltmeter
Table of Contents

Advertisement

871A & 871AB
3-7. 1 VOLT RANGE CALIBRATION
a.
Set switches on 871 as follows:
RANGE
1
NULL
od
voltage readout
1. 00000
dials
b.
Apply 1 volt DC +(0. 005% + 5 uv) between INPUT and COMMON posts.
c,
Adjust R121 for zero meter deflection in each successively more
sensitive null range.
3-8.
100 VOLT
RANGE
CALIBRATION
a.
Set switches on 871 as follows:
RANGE
100
NULL
10
voltage readout
100. 000
dials
b.
Apply 100 volts DC (+0. 00595) between
INPUT and COMMON
posts.
Note that calibration power supply must be readjusted for a galvanometer
null due to loading of voltmeter on voltage reference divider.
c.
Adjust R104 for zero meter deflection in each successively more
sensitive null range.
3-9.
1000 VOLT RANGE
CALIBRATION
a.
Set switches on 871 as follows:
RANGE
1. 000
NULL
100
voltage readout
1000. 00
dials
3-3