Respironics BiPAP autoSV Advanced User Manual page 42

With smartcard
Table of Contents

Advertisement

38
P
RESSURE
Output:
C
A
ONTROL
CCURACY
Parameter
Max Pressure
Min EPAP
Max EPAP
Min Pressure Support
Max Pressure Support
Breath Rate
Timed Inspiration
Ramp Duration
Rise Time
*
Dynamic pressure accuracy is ± 1.5 cm H
circuit with a Whisper Swivel II and varying fl ow conditions. Static pressure accuracy
is ± 1.25 cm H
and no patient fl ow.
** Th e range of values correspond to tenths of seconds (e.g., a setting of 4
indicates a Rise Time of 0.4 seconds).
*** Measured at the patient end of circuit with a Whisper Swivel II exhalation
device and no patient fl ow.
M
P
EASURED
ARAMETER
Parameter
Respiratory Rate
Exhaled Tidal Volume
Exhaled Minute
Ventilation
Leak Rate
WEEE/R
HS R
O
ECYCLING
If you are subject to the WEEE/RoHS recycling directives, refer to www.respironics.com for the
passport for recycling this product.
4 to 30 cm H
Range
4 to 30 cm
H
O
2
4 to 25 cm
H
O
2
4 to 25 cm
H
O
2
0 to 26 cm
H
O
2
0 to 26 cm
H
O
2
4 to 30 BPM
0.5 to 3.0 seconds
0 to 45 minutes
1 to 6 **
2
O measured at the patient end of the circuit with a Whisper Swivel II
2
A
CCURACY
Accuracy
Greater of ±1 BPM or ±10% of reading when
measured over a four minute period
± (25 + 0.15 of reading) ml
± (1 + 0.15 of reading) L/min
± (5 + 0.15 of reading) L/min
D
IRECTIVES
User Manual
O
2
Accuracy
*
*
*
*
*
Greater of ± 1 BPM or
± 10% of the setting (when
measured over a 4 minute
period)
± (0.1 + 10% of the setting)
seconds
± 10% of the setting
±
25%***
O measured at the patient end of the