Respironics BiPAP autoSV Advanced User Manual page 33

With smartcard
Table of Contents

Advertisement

7.3 A
T
LARM
ABLES
Th e following tables summarize the high priority, medium priority, and low priority alarms.
7.3.1 H
P
A
IGH
RIORITY
LARMS
Alarm
LED
• • • • •
Red Flash
• • • • •
Red Flash
• • • • •
Red Flash
• • • • •
Red Flash
• • • • •
Red Flash
Red Solid
Alarm
Display
Device
Message
Sound
Action
Operates
symbols
flash
Operates
symbols
flash
Operates
MinVent
symbols
flash
Shuts down.
symbol
Blower cannot
flashes
be restarted.
and an error
code ("Exx")
displays
Operates
and
cm H
O
2
symbols
flash
Blank
Shuts down
screen
User Manual
Possible
Your
Cause
Action
Press the
Breathing circuit is
RESET
RESET
button to reset the alarm.
disconnected or has
Reconnect the circuit
a large leak.
or fix the leak.
Press the
RESET
RESET
An apnea event
button to reset the alarm.
occurred during
Continue using the device.
therapy.
Report the alarm to your
health care professional.
Press the
A low minute
RESET
RESET
button to reset the
ventilation event
alarm. Continue using
occurred during
the device. Report the
therapy.
alarm to your health care
professional.
Press the
RESET
RESET
Device failure
button to reset the
alarm. Remove
power from the unit.
Restore power. If the
alarm continues,
contact your health care
professional.
Press the
RESET
RESET
Excessive leak
button to reset the
or blockage;
alarm.
malfunctioning unit.
Check for the
following: dirty inlet
filters, blocked air
intake, excessive leak
in the circuit. If the
alarm continues, call
your health care
professional.
Battery is discharged.
Press the
button to silence
the alarm.
-or-
Remove the DC power
source from the unit.
Replace the battery and
Power was lost
restore power to the
while the unit was
unit. Or, seek a reliable
providing therapy.
AC power source.
Restore power. If the
alarm continues,
call your health care
professional.