Respironics BiPAP autoSV Advanced User Manual page 34

With smartcard
Table of Contents

Advertisement

30
7.3.2 M
P
EDIUM
RIORITY
Alarm
Alarm
LED
Sound
• • •
Yellow Flash
DC Power
LED Flashes
7.3.3 L
P
A
OW
RIORITY
LARMS
Alarm
Alarm
LED
Sound
Yellow Solid
Yellow Solid
DC power
LED flashes
Yellow Solid
AC power
LED flashes
Yellow Solid
A
LARMS
Display
Device
Message
Action
Operates
Display
Device
Message
Action
• •
Operates
symbol
flashes
and
card error
code ("Cxx")
displays
• •
Operates
• •
Unchanged
Operates
• •
Operates
and cm H
O
2
symbols
flash
User Manual
Your
Possible
Action
Cause
Battery is nearly
Press the
RESET
discharged.
button to reset the
alarm. Replace the
battery. If the
alarm continues,
contact your health
care professional.
Possible
Your
Cause
Action
There is a problem
Confirm that the
with the SmartCard
SmartCard is properly
inserted in the
inserted.
SmartCard
connectivity slot.
If the alarm continues
to occur, remove the
Perhaps the
SmartCard from the
SmartCard is
device and contact
inserted upside down
your health care
or backwards.
professional.
Press the
The device lost AC
RESET
RESET
button to reset the
power and is now
alarm. Check the
operating on DC
power.
AC power. Seek a
reliable power
source. Provide AC
At start-up only,
power if you do not
alarm notifies you
want to use a battery;
that a battery is
otherwise, no further
being used
action is needed.
to provide power.
The AC power
Remove power from
supply is out of
the device and then
spec (<22V) or
restore power. If alarm
there is a defective
continues to occur,
battery sense line
contact your health
on the DC power
care professional.
adapter.
The device
Remove the
has successfully
SmartCard from the
downloaded the
device. If alarm
prescription from
continues to occur,
the SmartCard.
contact your health
care professional.