2.
Respiratory Rate Screen
Th is screen, shown in Figure 6–14, shows the average rate of respiration for the previous six breaths.
Th e display is updated at the end of each breath.
3.
Minute Ventilation Screen
Th is screen, shown in Figure 6–15, shows the estimated Exhaled Minute Ventilation (the volume of air
received on a per minute basis) based on the average of the previous six breaths.
NOTE:
Th e value shown for Exhaled Minute Ventilation is an estimate. Th e display fl ashes during
transient conditions such as low tidal volumes, erratic breathing, or rapidly changing leak.
4.
Exhaled Tidal Volume Screen
Th is screen, shown in Figure 6–16, shows the estimated Exhaled Tidal Volume, which is the volume
of each breath. Th e display is updated at the end of each breath.
NOTE:
Th e value shown for Exhaled Tidal Volume is an estimate. Th e display fl ashes during
transient conditions.
RR
Figure 6–14 Respiratory Rate Screen
MinVent
Figure 6–15 Minute Ventilation Screen
V
TE
Figure 6–16 Exhaled Tidal Volume Screen