Figure 34 Removing Display Assembly - Agilent Technologies InfiniiVision 7000 Series Service Manual

Hide thumbs Also See for InfiniiVision 7000 Series:
Table of Contents

Advertisement

5
Replacing Assemblies
98
9 Using a TORX T10 remove the four screws that secure the
display to the front deck. The upper right screw also
secures the inverter cable shield. Take care while
threading display cables thru front deck sheet metal
openings. Display cable clips/strain reliefs must be
disconnected prior to removing display assembly (see
figure 22).
Removing display assembly
Figure 34