Testing Performance
34
Equipment Required to Verify Voltage Measurement Accuracy
Table 6
Equipment
Critical Specifications
Oscilloscope
14 mV to 35 Vdc,
Calibrator
0.1 V resolution
Digital multimeter
Better than 0.01% accuracy
Cable
BNC, Qty 2
Shorting cap
BNC
Adapter
BNC (f) to banana (m)
Adapter
BNC tee (m) (f) (f)
Blocking capacitor
1 Set up the oscilloscope.
a Adjust the channel 1 position knob to place the
baseline at 0.5 major division from the bottom of the
display.
b Set the Volts/Div setting to the value in the first line in
Table
7.
Recommended
Model/Part
Fluke 5820A
Agilent 34401A
Agilent 10503A
Agilent 1250-0774
Agilent 1251-2277
Agilent 1250-0781 or
Pomona 3285
Agilent 10240-70001 +
Agilent 1250-0774 or
Agilent 11742A + Pomona
4288 + Pomona 5088A
7000 Series Oscilloscopes Service Guide