Oscilloscope
50
Feedthrough
N to BNC adapter
7000 Series Oscilloscopes Service Guide
1 Connect the equipment (see
a Use the N cable to connect the signal generator to the
power splitter input.
b Connect one output of the power splitter to the Aux
Trig input through a 50
c Connect the power sensor to the other output of the
power splitter.
Power Splitter
Power Sensor
Power Meter
Setting Up Equipment for 4-Channel External Trigger Sensitiv-
Figure 8
ity Test
Testing Performance
Figure
8).
feedthrough termination.
Signal
Generator
N Cable
2
55