To remove the AC input board

7000 Series Oscilloscopes Service Guide
1 Disconnect all cables on AC input board.
2 Disconnect the ground wire from its chassis terminal.
3 Locate and remove using a TORX T20 driver the single
screw securing the assembly to the rear deck.
4 Slide assembly to right and lift out of rear deck.
5 Take care that you do not damage the switch extender
during removal.
Removing the AC input board
Figure 39
Replacing Assemblies
Ground Wire
5