Test External Trigger Sensitivity (2-channel models)

7000 Series Oscilloscopes Service Guide
Verify the external trigger sensitivity at these settings:
Trigger range = +/- 1 V
500 MHz (MSO/DSO7052A/B), 200 mV
350 MHz (MSO/DSO7032A/B), 200 mV
100 MHz (MSO/DSO7012A/B), 200 mV
Trigger range = +/- 8 V
500 MHz (MSO/DSO7052A/B), 500 mV
350 MHz (MSO/DSO7032A/B), 500 mV
100 MHz (MSO/DSO7012A/B), 500 mV
Testing Performance