2
Testing Performance

Overview

22
To completely test and troubleshoot MSO models, create and
use the test connector accessory as described in this
chapter.
• The test connector is only required for oscilloscopes that
have the MSO option licensed (enabled).
• The connector is used in the digital channel threshold
accuracy test.
• The test connector keeps electrical distortion to a
minimum and makes it easy for you to connect the
oscilloscope probes to function generators and
measurement equipment.
Let the Equipment Warm Up Before Testing
For accurate test results, let the test equipment and the
oscilloscope warm up 30 minutes before testing.
Verifying Test Results
During the tests, record the readings in the Performance
Test Record on
page
verify that the reading is within the limits in the
Performance Test Record.
If a performance test fails
If a performance test fails, first perform the User Cal procedure. Press the
following keys to access User Cal: Utility Service Start User Cal.
57. To verify whether a test passes,
7000 Series Oscilloscopes Service Guide