2
Testing Performance
52
1 Connect the equipment (see
a Use the N cable to connect the signal generator to the
power splitter input.
b Connect one output of the power splitter to the Ext
Trigger input.
c Connect the power sensor to the other output of the
power splitter.
Oscilloscope
50
Feedthrough
N to BNC adapter
Power Sensor
Figure 7
Power Splitter
N Cable
Power Meter
Setting Up Equipment for 2-Channel External Trigger Sensitiv-
ity Test
7000 Series Oscilloscopes Service Guide
Figure
7).
Signal
Generator