Calibrating And Adjusting - Agilent Technologies InfiniiVision 7000 Series Service Manual

Hide thumbs Also See for InfiniiVision 7000 Series:
Table of Contents

Advertisement

Agilent InfiniiVision 7000 Series Oscilloscope
Service Guide
3

Calibrating and Adjusting

This chapter explains how to adjust the oscilloscope for
optimum operating performance. You should perform
self- calibration according to the following recommendations:
• Every 12 months or after 2000 hours of operation
• If the ambient temperature is >10 °C from the calibration
temperature
• If you want to maximize the measurement accuracy
The amount of use, environmental conditions, and
experience with other instruments help determine if you
need shorter adjustment intervals.
Let the Equipment Warm Up Before Adjusting
Before you start the adjustments, let the oscilloscope and
test equipment warm up for at least 30 minutes.
Read All Cautions and Warnings
Read the following cautions and warning before making
adjustments or performing self- calibration.
HAZARDOUS VOLTAGES!
WA R N I N G
Read the safety notice at the front of this book before proceeding.
Maintenance is performed with power supplied to the oscilloscope and with the
protective covers removed. Only trained service personnel who are aware of the
hazards involved should perform the maintenance. Whenever possible, perform the
procedures with the power cord removed from the oscilloscope.
REMOVE POWER TO AVOID DAMAGE!
C A U T I O N
Do not disconnect any cables or remove any assemblies with power applied to the
oscilloscope. Otherwise, damage to the oscilloscope can occur.
s1
59

Advertisement

Table of Contents
loading

Table of Contents