7000 Series Oscilloscopes Service Guide
BNC Tee
Test
Connector
Setting Up Equipment for Digital Channel Threshold Accuracy
Figure 2
Test
4 Use a BNC- banana cable to connect the multimeter to the
other side of the BNC Tee.
5 Connect the BNC Tee to the Channel 1 output of the
calibrator as shown in
6 On the oscilloscope, press the D15-D0 key, then press the
Thresholds softkey, then press the D7 - D0 softkey repeatedly
until the check mark is next to User.
7 Press the User softkey to the right of the D7 - D0 softkey,
then turn the Entry knob (
Testing Performance
7000 Series Oscilloscope
Channels
8 - 15
Channels
0 - 7
Figure
2.
) on the front panel of the
2
Probe
Cables
31