Table 8 Equipment Required To Verify Bandwidth - Agilent Technologies InfiniiVision 7000 Series Service Manual

Hide thumbs Also See for InfiniiVision 7000 Series:
Table of Contents

Advertisement

7000 Series Oscilloscopes Service Guide
Equipment Required to Verify Bandwidth
Table 8
Equipment
Critical Specifications
Signal Generator
100 kHz - 1 GHz at 200 mVrms Agilent N5181A,
Power Meter/Sensor
1 MHz - 1 GHz ±3% accuracy
Power Splitter
outputs differ by < 0.15 dB
Cable
Type N (m) 24 inch
Adapter
Type N (m) to BNC (m)
1 Connect the equipment (see
a Use the N cable to connect the signal generator to the
input of the power splitter input.
b Connect the power sensor to one output of the power
splitter.
c Use an N- to- BNC adapter to connect the other splitter
output to the channel 1 input.
Testing Performance
Recommended
Model/Part
E4400B, or 8648A
Agilent E4418B/8482A
Agilent 11667A
Agilent 11500B
Agilent 1250-0082 or
Pomona 3288 with
Pomona 3533
Figure
5).
2