Viewing and Measuring Digital Signals

To connect the digital probes to the circuit under test

C A U T I O N
76
1 If you feel it's necessary, turn off the power supply to the
circuit under test.
Turning off power to the circuit under test would only
prevent damage that might occur if you accidentally short
two lines together while connecting probes. You can leave the
oscilloscope powered on because no voltage appears at the
probes.
2 Connect the digital probe cable to the D15 - D0 connector on
the rear panel of the mixed-signal oscilloscope. The digital
probe cable is indexed so you can connect it only one way.
You do not need to power-off the oscilloscope.
3 Route the cable under the oscilloscope and bring it out in
front. Lay the two sections of the digital cable on top of each
other at the front of the oscilloscope. Slip the narrow end of
the cable guide over the cable and into its slots at the front of
the oscilloscope. Snap the wide end of the cable guide in
place, being careful not to pinch the cable.
Use only the Agilent part number 54620-68701 digital probe kit
supplied with the mixed-signal oscilloscope.