Agilent Technologies 6000 Series User Manual page 97

Oscilloscope
Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

C A U T I O N
C A U T I O N
C A U T I O N
6000 Series Oscilloscope User's Guide
Probe Attenuation
Turn the Entry knob to set the attenuation
factor displayed in the Probe softkey for the connected probe.
The attenuation factor can be set from 0.1:1 to 1000:1 in a 1-2-5
sequence.
When you connect an AutoProbe self-sensing probe, the
oscilloscope will automatically configure your probe to the
correct attenuation factor. Note that the AutoProbe feature is
not available on 100 MHz bandwidth models.
The probe correction factor must be set properly for
measurements to be made correctly.
Range
The input voltage range can be set to 1.0 Volts or 8.0
Volts. When in current mode, the range is fixed at 1.0 Amps.
Maximum input voltage for the external trigger input of the
2-channel oscilloscope:
Maximum input voltage for analog inputs:
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 100 Vrms, 400 Vpk
with 10073C or 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
Do not exceed 5 Vrms in 50
protection is enabled in 50
greater than 5 Vrms is detected. However the input could still be
damaged, depending on the time constant of the signal.
The 50
input protection mode only functions when the oscilloscope
is powered on.
Input Impedance
The 300 MHz, 500 MHz, and 1 GHz bandwidth
2-channel oscilloscopes have selectable external trigger input
impedance. The impedance can be set to either 1M Ohm or
50 Ohm by pressing the Imped softkey. The input impedance of
100 MHz bandwidth models is fixed at 1M .
Triggering the Oscilloscope
mode on the 2-channel models. Input
mode and the 50
load will disconnect if
4
97

Advertisement

Table of Contents
loading

Table of Contents