To restore the oscilloscope to its default configuration

6000 Series Oscilloscope User's Guide
Installed licenses:
This line in the About This Oscilloscope dialog contains
information about the licenses that have been installed on the
oscilloscope. For example, it can show:
• MSO — Mixed Signal Oscilloscope. This feature adds 16 digital
channels without having to install any hardware.
• LSS — Low Speed Serial Decode. This license provides serial
2
decode of I
C and SPI buses.
• AMS — Automotive Serial Decode. This license provides serial
decode of CAN and LIN buses.
• memXM — Memory Upgrade. This license provides additional
memory, increasing the total memory depth to 2 MB or 8 MB
(interleaved).
• Press the Save/Recall key, then press the Default Setup softkey.
The default configuration returns the oscilloscope to its default
settings. This places the oscilloscope in a known operating
condition. The major default settings are:
Horizontal
main mode, 100 s/div scale, 0 s delay, center time
reference.
Vertical (Analog)
Channel 1 on, 5 V/div scale, DC coupling, 0 V
position, 1 M impedance, probe factor to 1.0 if an AutoProbe
probe is not connected to the channel.
Trigger
Edge trigger, Auto sweep mode, 0 V level, channel 1
source, DC coupling, rising edge slope, 60 ns holdoff time.
Display
Vectors on, 33% grid intensity, infinite persistence off.
Other
Acquire mode normal, Run/Stop to Run, cursors and
measurements off.
Front-Panel Controls
2
73