Displaying Data
226
4 Specify the CAN signal:
a Press the Settings softkey to access the CAN Signals Menu.
Baud rate
Signal
Source
5 Press the Source softkey, and use the Entry knob to select the
channel for the CAN signal. Make sure the trigger or
threshold levels are set to the middle of the CAN signal.
6 Repeatedly press and release the Baud softkey to specify the
CAN signal baud rate.
7 Repeatedly press and release the Smpl Pt softkey to select the
point between phase segments 1 and 2 where the state of the
Currently selected
Sample
Point
6000 Series Oscilloscope User's Guide
Trigger level
channel
Return to
previous menu