Agilent Technologies 6000 Series User Manual page 174

Oscilloscope
Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

5
Making Measurements
174
The following FFT spectrum was obtained by connecting the
front panel Probe Comp signal (~1.2 kHz) to channel 1. Set
sweep speed to 5 ms/div, vertical sensitivity to 500 mV/div,
Units/div to 10 dBV, Offset to -34.0 dBV, Center frequency to
5.00 kHz, frequency Span to 10.0 kHz, and window to Hanning.
Figure 27
FFT measurements
FFT Measurement Hints
The number of points acquired for the FFT record is 1000 and
when frequency span is at maximum, all points are displayed.
Once the FFT spectrum is displayed, the frequency span and
center frequency controls are used much like the controls of a
spectrum analyzer to examine the frequency of interest in
greater detail. Place the desired part of the waveform at the
center of the screen and decrease frequency span to increase
the display resolution. As frequency span is decreased, the
number of points shown is reduced, and the display is
magnified.
While the FFT spectrum is displayed, use the Math and Cursors
keys to switch between measurement functions and frequency
domain controls in FFT menu.
6000 Series Oscilloscope User's Guide

Advertisement

Table of Contents
loading

Table of Contents