Making Measurements
200
Overshoot
Overshoot is distortion that follows a major edge transition
expressed as a percentage of Amplitude. The X cursors show
which edge is being measured (edge closest to the trigger
reference point).
Rising edge overshoot =
Falling edge overshoot =
Overshoot
local Minimum
Figure 36
Automatic Overshoot measurement
local Maximum - D Top
--------------------------------------------------------- - X 100
Amplitude
Base - D local Minimum
---------------------------------------------------------- - X 100
Amplitude
local Maximum
Overshoot
6000 Series Oscilloscope User's Guide
Top
Base