Calibration Process - Campbell CR510 Operator's Manual

Datalogger
Hide thumbs Also See for CR510:
Table of Contents

Advertisement

In Figure 13.5-2, V
x
is a fixed resistor, R
and R
is the resistance between the excited
G
electrode and CR510 earth ground. With R
the network, the measured signal is:
V
= V
__________________
1
x
(R
s
R
R
/R
is the source of error due to the ground
s
f
G
loop. When R
is large the equation reduces to
G
the ideal. The geometry of the electrodes has a
great effect on the magnitude of this error. The
Delmhorst gypsum block used in the 227 probe
has two concentric cylindrical electrodes. The
center electrode is used for excitation; because
it is encircled by the ground electrode, the path
for a ground loop through the soil is greatly
reduced. Moisture blocks which consist of two
parallel plate electrodes are particularly
susceptible to ground loop problems. Similar
considerations apply to the geometry of the
electrodes in water conductivity sensors.
The ground electrode of the conductivity or soil
moisture probe and the CR510 earth ground
form a galvanic cell, with the water/soil solution
acting as the electrolyte. If current was allowed
to flow, the resulting oxidation or reduction
would soon damage the electrode, just as if DC
excitation was used to make the measurement.
Campbell Scientific probes are built with series
capacitors in the leads to block this DC current.
In addition to preventing sensor deterioration,
the capacitors block any DC component from
affecting the measurement.

13.6 CALIBRATION PROCESS

The CR510 makes voltage measurements by
integrating the input signal for a fixed time and
then holding the integrated value for the analog
to digital (A/D) conversion. The A/D conversion
is made by a 13 bit approximation using a digital
to analog converter (DAC). The result from the
approximation is DAC counts, which are
multiplied by coefficients to obtain millivolts
(mV). There are 10 calibration coefficients, one
for each of the 5 gain ranges for the fast and
slow integration times.
The CR510 has an internal calibration function
that feeds positive and negative voltages
through the amplifiers and integrator and
calculates new calibration coefficients. By
is the excitation voltage, R
is the sensor resistance,
s
R
s
[13.6-1]
+R
) + R
R
/R
f
s
f
G
SECTION 13. CR510 MEASUREMENTS
adjusting the calibration coefficients the
f
accuracy of the voltage measurements is
maintained over the -25 to +50 C operating
in
range of the CR510. Calibration is executed
G
under four conditions:
1. When the CR510 is powered up.
2. Automatically when Instruction 24 is not
contained in a program table.
3. When the watchdog resets the processor.
4. When the calibration instruction, Instruction
24, is executed.
AUTOMATIC CALIBRATION SEQUENCE
The primary advantage of automatic calibration
is that the CR510 is constantly calibrated
without user programming. The CR510 defaults
to automatic calibration when Instruction 24 is
not contained in a program table.
Every 8 seconds one part of a 22 part
calibration sequence is performed. Program
execution is interrupted (5.4 - 21.4 ms), when
necessary, for each part of the calibration.
Every 2.9 minutes (8 seconds * 22) ten
calibration coefficients are calculated. The
calculated coefficients are multiplied by 1/5, and
then added to 4/5 times the existing
coefficients. Averaging is done as a safeguard
against coefficients calculated from a noisy
measurement.
The above weighting of the newly calculated
coefficients results in a 15 minute time constant
(see Instruction 58) in the response of the
calibration to step changes affecting the
calibration coefficients (primarily temperature).
For most environmental applications a 15
minute time constant is acceptable. The
automatic calibration may result in the
calibration coefficients not being optimum for
applications that subject the CR510 to extreme
temperature gradients.
Automatic calibration extends the processing
time 5.4 to 21.4 ms when it is executed (every 8
seconds). If the processing time exceeds the
execution interval the CR510 finishes
processing the table and awaits the next
occurrence of the execution interval before
initiating the table. At the fastest execution
interval of 1/64 (0.0156) second the program
13-17

Advertisement

Table of Contents
loading

Table of Contents