The Effect Of Sensor Lead Length On The Signal Settling Time - Campbell CR510 Operator's Manual

Datalogger
Hide thumbs Also See for CR510:
Table of Contents

Advertisement

which both inputs of a differential measurement
must lie in order for the differential measurement
to be made.
For example, if the high side of a differential input
is at 2 V and the low side is at 1 V relative to
CR510 ground, there is no problem; a
measurement made on the +2.5 V range would
indicate a signal of 1 V. However, if the high input
is at 2.8 V and the low input is at 2 V, the
measurement cannot be made because the high
input is outside of the common mode range. The
CR510 will indicate the overrange with the
maximum negative number (Section 3.5.)
Problems with exceeding common mode range
may be encountered when the CR510 is used
to read the output of external signal conditioning
circuitry if a good ground connection does not
exist between the external circuitry and the
CR510. When operating where AC power is
available, it is not always safe to assume that a
good ground connection exists through the AC
wiring. If a CR510 is used to measure the
output from a laboratory instrument (both
plugged into AC power and referencing ground
to outlet ground), it is best to run a ground wire
between the CR510 and the external circuitry.
Even with this ground connection, the ground
potential of the two instruments may not be at
exactly the same level, which is why a
differential measurement is desired (Section
7.2).
If a differential measurement is used on a sensor
that is not referenced to CR510 ground through a
separate connection (e.g., a net radiometer), a
jumper wire should be connected between the low
side of the differential input and analog ground to
hold the sensor in common mode range.
A differential measurement has better noise
rejection than a single-ended measurement.
Integrating the signal in both directions also
reduces input offset voltage due to thermal
effects in the amplifier section of the CR510.
Input offset voltage on a single-ended
measurement is less than 5 microvolts; the
input offset voltage on a differential
measurement is less than 1 microvolt.
A single-ended measurement is quite
satisfactory in cases where noise is not a
problem and care is taken to avoid ground
potential problems. Channels are available for
SECTION 13. CR510 MEASUREMENTS
twice as many single-ended measurements. A
single-ended measurement takes about half the
time of a differential measurement, which is
valuable in cases where rapid sampling is a
requirement.
NOTE: Sustained voltages in excess of
+16 VDC applied to the analog inputs will
damage the CR510 input circuitry.
13.3 THE EFFECT OF SENSOR LEAD
LENGTH ON THE SIGNAL SETTLING
TIME
Whenever an analog input is switched into the
CR510 measurement circuitry prior to making a
measurement, a finite amount of time is
required for the signal to stabilize at its correct
value. The rate at which the signal settles is
determined by the input settling time constant
which is a function of both the source
resistance, and input capacitance (explained
below). The CR510 allows a 450 µs settling
time before initiating the measurement. In most
applications this settling time is adequate, but
the additional wire capacitance associated with
long sensor leads can increase the settling time
constant to the point that measurement errors
may occur. There are three potential sources of
error which must settle before the measurement
is made:
1. The signal must rise to its correct value.
2. A small transient (~5 mV) caused by
switching the analog input into the
measurement circuitry must settle.
3. A larger transient, usually about 40 mV/V,
caused by the switched, precision excitation
voltage used in resistive bridge
measurements must settle.
The purpose of this section is to bring attention
to potential measurement errors caused when
the input settling time constant gets too large
and to discuss procedures whereby the effects
of lead length on the measurement can be
estimated. In addition, physical values are given
for three types of wire used in CSI sensors, and
error estimates for given lead lengths are
provided. Finally, techniques are discussed for
minimizing input settling error when long leads
are mandatory.
13-3

Advertisement

Table of Contents
loading

Table of Contents